Evaluation of RF Heating on Humerus Implant in Phantoms during 1.5T MR Imaging and Comparisons with Electromagnetic Simulation

https://hirokoku-u.repo.nii.ac.jp/records/943
https://hirokoku-u.repo.nii.ac.jp/records/943
e72615da-f231-44c6-9855-c1b5b5022683
名前 / ファイル ライセンス アクション
MRMS-5(2).pdf MRMS-5(2).pdf (2.7 MB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2023-03-17
タイトル
タイトル Evaluation of RF Heating on Humerus Implant in Phantoms during 1.5T MR Imaging and Comparisons with Electromagnetic Simulation
言語 en
言語
言語 eng
キーワード
主題Scheme Other
主題 magnetic resonance imaging
キーワード
主題Scheme Other
主題 metallic implant
キーワード
主題Scheme Other
主題 RF heating
キーワード
主題Scheme Other
主題 electromagnetic simulation
キーワード
主題Scheme Other
主題 MRI safety
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 村中, 博幸

× 村中, 博幸

ja 村中, 博幸

ja-Kana ムラナカ, ヒロユキ

Search repository
堀口, 隆良

× 堀口, 隆良

ja 堀口, 隆良

ja-Kana ホリグチ, タカヨシ

Search repository
笛吹, 修治

× 笛吹, 修治

ja 笛吹, 修治

ja-Kana ウスイ, シュウジ

Search repository
上田, 善武

× 上田, 善武

ja 上田, 善武

ja-Kana ウエダ, ヨシタケ

Search repository
中村, 修

× 中村, 修

ja 中村, 修

ja-Kana ナカムラ, オサム

Search repository
池田, 文昭

× 池田, 文昭

ja 池田, 文昭

ja-Kana イケダ, フミアキ

Search repository
岩倉, 健

× 岩倉, 健

ja 岩倉, 健

ja-Kana イワクラ, ケン

Search repository
中谷, 儀一郎

× 中谷, 儀一郎

ja 中谷, 儀一郎

ja-Kana ナカヤ, ギイチロウ

Search repository
著者(英)
姓名 Muranaka, Hiroyuki
言語 en
著者(英)
姓名 Horiguchi, Takayoshi
言語 en
著者(英)
姓名 Usui, Shuji
言語 en
著者(英)
姓名 Ueda, Yoshitake
言語 en
著者(英)
姓名 Nakamura, Osamu
言語 en
著者(英)
姓名 Ikeda, Fumiaki
言語 en
著者(英)
姓名 Iwakura, Ken
言語 en
著者(英)
姓名 Nakaya, Giichirou
言語 en
抄録
内容記述タイプ Abstract
内容記述 Purpose: To evaluate the effect of radiofrequency (RF) heating on a metallic implant during magnetic resonance imaging (MRI), temperatures at several positions of an implant were measured, and results are compared with electromagnetic simulations using a finite element method.Methods: A humerus nail implant made of stainless steel was embedded at various depths of tissue-equivalent gel-phantoms with loop (loop phantom) and partially cut loop (loop-cut phantom), and the phantoms were placed parallel to the static magnetic field of a 1.5T MRI device. Scans were conducted at maximum RF for 15 min, and temperatures were recorded with 2 RF-transparent fiberoptic sensors. Finally, electromagnetic-field analysis was performed.Results: Temperatures increased at both ends of the implants at various depths, and temperature increase was suppressed with increasing depth. The maximum temperature rise was 12.3°C at the tip of the implant and decreased for the loop-cut phantom. These tendencies resembled the results of electromagnetic simulations.Conclusion: RF heating was verified even in a nonmagnetizing metal implant in a case of excessive RF irradiation. Particularly, rapid temperature rise was observed at both ends of the implant having large curvatures. The difference in temperature increase by depth was found to reflect the skin-depth effect of RF intensity. Electromagnetic simulation was extremely useful for visualizing the eddy currents within the loop and loop-cut phantoms and for evaluating RF heating of a metallic implant for MRI safety.
言語 en
書誌情報 Magnetic resonance in medical sciences : MRMS

巻 5, 号 2, p. 79-88, 発行日 2006
出版者
出版者 日本磁気共鳴医学会
ISSN
収録物識別子タイプ PISSN
収録物識別子 1347-3182
書誌レコードID
収録物識別子タイプ NCID
収録物識別子 AA11648770
権利
権利情報 本論文の著作権は日本磁気共鳴医学会に属する
著者版フラグ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-25 10:39:13.030736
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3