定位放射線治療における品質管理サポートシステムの構築
https://doi.org/10.60172/00000341
https://doi.org/10.60172/00000341f27ab5b2-40fa-45e7-995e-55f08055b104
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 紀要論文 / Departmental Bulletin Paper(1) | |||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
公開日 | 2023-03-11 | |||||||||||||||||
タイトル | ||||||||||||||||||
タイトル | 定位放射線治療における品質管理サポートシステムの構築 | |||||||||||||||||
言語 | ja | |||||||||||||||||
タイトル | ||||||||||||||||||
タイトル | Development of quality control support system for stereotactic radiation therapy | |||||||||||||||||
言語 | en | |||||||||||||||||
言語 | ||||||||||||||||||
言語 | jpn | |||||||||||||||||
キーワード | ||||||||||||||||||
主題Scheme | Other | |||||||||||||||||
主題 | 定位放射線治療 | |||||||||||||||||
キーワード | ||||||||||||||||||
主題Scheme | Other | |||||||||||||||||
主題 | 画像処理 | |||||||||||||||||
キーワード | ||||||||||||||||||
主題Scheme | Other | |||||||||||||||||
主題 | 品質保証 | |||||||||||||||||
キーワード | ||||||||||||||||||
主題Scheme | Other | |||||||||||||||||
主題 | 品質管理 | |||||||||||||||||
資源タイプ | ||||||||||||||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||||||||||||||
資源タイプ | departmental bulletin paper | |||||||||||||||||
ID登録 | ||||||||||||||||||
ID登録 | 10.60172/00000341 | |||||||||||||||||
ID登録タイプ | JaLC | |||||||||||||||||
著者 |
田邉, 悦章
× 田邉, 悦章
× 大倉, 保彦
× 石田, 隆行
|
|||||||||||||||||
著者(英) | ||||||||||||||||||
姓名 | Tanabe, Yoshiori | |||||||||||||||||
言語 | en | |||||||||||||||||
著者(英) | ||||||||||||||||||
姓名 | Okura, Yasuhiko | |||||||||||||||||
言語 | en | |||||||||||||||||
著者(英) | ||||||||||||||||||
姓名 | Ishida, Takayuki | |||||||||||||||||
言語 | en | |||||||||||||||||
抄録 | ||||||||||||||||||
内容記述タイプ | Abstract | |||||||||||||||||
内容記述 | Stereotactic radio therapy (SRT) using a linear Accelerator (Linac) requires high precision in the table position and the center of irradiation. QC procedure is necessary To guarantee high precision of each SRT. As the QC procedure, we use the film shot method, in which the table correction width is derived from target images taken at various gantry angles and table angles. However, QC procedure is so complex that long time is wasted for this procedure. In addition, the degree of precision of QC procedure depends on each technologist. We developed a QC program support system to support QC procedure. This system processes images taken by the film shot method, and outputs the table correction width to adjust target position. For evaluation of our developed system, we compared correction width of obtained from our system with that from, conventional manual operation. Results showed that high degree of precision was achieved by employing our system. By using our system, one double check scheme, QC procedure by using both manual operation and our system, will be realized without any additional time for the procedure. | |||||||||||||||||
言語 | en | |||||||||||||||||
書誌情報 |
医療工学雑誌 en : Journal of Medical Engineering 号 1, p. 45-49, 発行日 2007-03-15 |
|||||||||||||||||
出版者 | ||||||||||||||||||
出版者 | 広島国際大学大学院総合人間科学研究科医療工学専攻 | |||||||||||||||||
ISSN | ||||||||||||||||||
収録物識別子タイプ | PISSN | |||||||||||||||||
収録物識別子 | 1881-9265 | |||||||||||||||||
書誌レコードID | ||||||||||||||||||
収録物識別子タイプ | NCID | |||||||||||||||||
収録物識別子 | AA12217173 | |||||||||||||||||
著者版フラグ | ||||||||||||||||||
出版タイプ | VoR | |||||||||||||||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 |