タイトル | 透析技術教育用シミュレータのためのplasma refilling rate 簡易推定法の検討 | |||||||||||||
言語 | ja | |||||||||||||
タイトル | ||||||||||||||
タイトル | Examination of the simple estimation method of the plasma refilling rete for the hemodialysis training simulator system | |||||||||||||
言語 | en | |||||||||||||
言語 | ||||||||||||||
言語 | jpn | |||||||||||||
キーワード | ||||||||||||||
主題Scheme | Other | |||||||||||||
主題 | plasma refilling rate | |||||||||||||
キーワード | ||||||||||||||
主題Scheme | Other | |||||||||||||
主題 | hematocrit sensor | |||||||||||||
キーワード | ||||||||||||||
主題Scheme | Other | |||||||||||||
主題 | total blood volume | |||||||||||||
キーワード | ||||||||||||||
主題Scheme | Other | |||||||||||||
主題 | patient simulator | |||||||||||||
キーワード | ||||||||||||||
主題Scheme | Other | |||||||||||||
主題 | hemodialysis | |||||||||||||
資源タイプ | ||||||||||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||||||||||
資源タイプ | departmental bulletin paper | |||||||||||||
ID登録 | ||||||||||||||
ID登録 | 10.60172/00000321 | |||||||||||||
ID登録タイプ | JaLC | |||||||||||||
著者 |
戸梶, めぐみ
× 戸梶, めぐみ
× 二宮, 伸治
|
|||||||||||||
著者(英) | ||||||||||||||
姓名 | Tokaji, Megumi | |||||||||||||
言語 | en | |||||||||||||
著者(英) | ||||||||||||||
姓名 | Ninomiya, Shinji | |||||||||||||
言語 | en | |||||||||||||
抄録 | ||||||||||||||
内容記述タイプ | Abstract | |||||||||||||
内容記述 | Because the numbers of hemodialysis patients are increasing, it is important to establish the educational procedure for clinical engineers to keep the safety of hemodialysis procedure. Therefore the development of the high-fidelity hemodialysis simulation system which can use for education of hemodialysis techniques is expected. To simulate the hemodynamic response of patients, it is necessary to develop the numerical model of patient’s condition during the hemodialysis. However, since the change of plasma refilling rate (PRR) and baroreflex control of the cardiovascular system affect the hemodynamics under dialysis intricately, construction of a correctly reproducible scientific model is very difficult. In this study, we investigated the simple estimation method to define the PRR from the patient’s initial conditions. We evaluated the PRR and circulating blood volume change during hemodialysis using CRIT-LINE in 17 patients under the constant ultrafiltration rate (UFR). Time-averaged PRR/UFR value was obtained as 0.8215±0.10 under the different total amount of ultra-filtration rate(UFR) and total blood volume. Hence we found the PRR value can estimate from UFR directly by using the proper PRR/UFR coefficient. This simple estimation method seems to be effective to construct the practical model which can simulate the hemodynamic response under hemodialysis. | |||||||||||||
言語 | en | |||||||||||||
書誌情報 |
医療工学雑誌 en : Journal of Medical Engineering 号 3, p. 1-5, 発行日 2009-03-15 |
|||||||||||||
出版者 | ||||||||||||||
出版者 | 広島国際大学大学院総合人間科学研究科医療工学専攻 | |||||||||||||
ISSN | ||||||||||||||
収録物識別子タイプ | PISSN | |||||||||||||
収録物識別子 | 1881-9265 | |||||||||||||
書誌レコードID | ||||||||||||||
収録物識別子タイプ | NCID | |||||||||||||
収録物識別子 | AA12217173 | |||||||||||||
著者版フラグ | ||||||||||||||
出版タイプ | VoR | |||||||||||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 |