Ex Vivo Coagulation Test on Tissue Factor-Expressing Cells with a Calibrated Automated Thrombogram

https://hirokoku-u.repo.nii.ac.jp/records/1043
https://hirokoku-u.repo.nii.ac.jp/records/1043
ba6a932c-f467-44ee-9d5d-c93e7555a6e6
名前 / ファイル ライセンス アクション
039-044.08010.Takamiya.pdf 039-044.08010.Takamiya.pdf (5.0 MB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2023-03-17
タイトル
タイトル Ex Vivo Coagulation Test on Tissue Factor-Expressing Cells with a Calibrated Automated Thrombogram
言語 en
言語
言語 eng
キーワード
主題Scheme Other
主題 Ex vivo test
キーワード
主題Scheme Other
主題 Thrombin generation
キーワード
主題Scheme Other
主題 Tissue factor-bearing cell
キーワード
主題Scheme Other
主題 Coagulation factor-deficient plasma
キーワード
主題Scheme Other
主題 Calibrated automated thrombography
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 高宮, 脩

× 高宮, 脩

ja 高宮, 脩

ja-Kana タカミヤ, オサム

Search repository
阪田, 光彦

× 阪田, 光彦

ja 阪田, 光彦

ja-Kana サカタ, ミツヒコ

Search repository
著者(英)
姓名 Takamiya, Osamu
言語 en
著者(英)
姓名 Sakata, Mitsuhiko
言語 en
抄録
内容記述タイプ Abstract
内容記述 A calibrated automated thrombogram is not affected by the turbidity of platelet and cell preparations because the measurement is based on fluorescence. To examine conditions that mimic the physiological state, we investigated thrombograms that show thrombin generation on tissue factor (TF)-bearing cells. An increase in the number of J82 cells did not affect the endogenous thrombin potential (ETP) of normal plasma, although the lag time (LT), the peak height, and the time to peak (ttPeak) did depend on cell concentration. When 5 parameters of coagulation factor-deficient plasmas were plotted on a radar graph, the thrombogram pattern of factor XI (FXI)-deficient plasma became slightly reduced. The thrombogram did not improve when washed normal platelets or washed normal platelets with adenosine diphosphate (ADP) were added. FVII-depleted plasma, FVIII-deficient plasma, and FIX-deficient plasma showed remarkably reduced peak heights, ttPeaks, and times to the end of thrombin generation (start tails). The thrombogram of FVII-depleted plasma was characterized by a remarkably prolonged LT, unlike the patterns of FVIII- or FIX-deficient plasma and FXI-depleted plasma. The ETP of FVIII- and FIX-deficient plasma, but not FVII-depleted plasma, improved significantly upon addition of washed normal platelets or washed normal platelets with ADP. The thrombograms of coagulation factor-deficient plasma containing TF-bearing cells differed from those for recombinant TF and phospholipid in the liquid phase. We suggest that thrombograms using TF-bearing cells can be a useful ex vivo test, because this experimental model may be analogous to most coagulation processes in vivo.
言語 en
内容記述
内容記述タイプ Other
内容記述 論文に記載されている著者名は O.TAKAMIYA, M.SAKATA
書誌情報 Laboratory hematology : official publication of the International Society for Laboratory Hematology.

巻 14, 号 4, p. 39-44, 発行日 2008
出版者
出版者 Kluge Carden Jennings Pub. Co.
ISSN
収録物識別子タイプ PISSN
収録物識別子 1080-2924
書誌レコードID
収録物識別子タイプ NCID
収録物識別子 AA11859936
著者版フラグ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-25 10:38:57.079310
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3